
User Manual
i Series / iX Series 321
Command could not
be executed
Command may be inconsistent with
mode of operation such as
programming frequency when in DC
mode.
Command issued but
unit is not in remote
state
Put instrument in remote state before
issuing GPIB commands.
Some commands are supported by
the unit but are locked out for
protection of settings and are not
user accessible.
Problem with trigger
system.
Unit could not generate trigger for
transient execution or measurement.
Trigger request has
been ignored.
Trigger setup incorrect or unit was not
armed when trigger was received.
Check transient system or
measurement trigger system settings.
Initialization request
has been ignored
Unit was told to go to armed state but
was unable to do so. Could be
caused by incorrect transient system
or measurement acquisition setup.
Incorrect parameter or parameter
value. Check programming manual
for allowable parameters
Requested setting
conflicts with other
setting in effect.
Check other settings. E.g. trying to
program a DC offset while in AC
mode
Parameter data
outside of allowable
range.
Check programming manual for
allowable parameter values
More data received
than expected
Check programming manual for
number of parameters or data block
size
"Illegal parameter
value"
Parameter value is
not suppored
Check programming manual for
correct parameters
One or more
transient lists
programmed has
different length.
All lists must be of same length or
transient cannot be compiled and
executed.
No storage space left
to save settings or
data.
Delete other settings or data to make
room.
Too many waveform
directory entries
Delete one or more waveforms from
waveform memory to make room.
Waveform requested
not in directory
Check waveform directory for
waveform names present.
Too many or non ASCII characters
used in waveform file definition.
Use ASCII characters only
Check hardware for proper operation.
Comentarios a estos manuales